Title Artificial Intelligence on a Smartphone Versus the Ruler: Objective Measurement of Ptosis Severity
Number 302
Author Christine Safieh
Principal Unit where research was undertaken MID Yorkshire Teaching NHS Trust

Purpose Accurate measurement of marginal reflex distance (MRD1) is essential in the assessment and management of ptosis. Traditional ruler-based measurements are widely used but may be subject to inter-observer variability and measurement error. Smartphone-based artificial intelligence (AI) tools may offer a rapid and objective alternative. This study aimed to compare AI-assisted smartphone measurement of MRD1 with conventional ruler-based clinical assessment.
Methods A prospective observational study was conducted in patients presenting with upper eyelid ptosis. Standardised frontal photographs were captured using a smartphone camera under consistent clinical conditions. MRD1 was measured manually using a millimetre ruler by two independent clinicians. The same images were analysed using a smartphone-based AI facial landmark detection algorithm to automatically calculate MRD1. Agreement between AI-derived and clinician-measured MRD1 values was evaluated using Pearson correlation and Bland–Altman analysis. Measurement time and inter-observer variability were also recorded.
Results Sixty-two patients (124 eyelids) were included in the analysis. AI-derived MRD1 measurements showed strong correlation with manual ruler measurements (r = 0.90, p < 0.001). Bland–Altman analysis demonstrated a mean difference of 0.06 mm with 95% limits of agreement between −0.41 mm and 0.53 mm. Inter-observer variability was reduced using the AI method compared with manual measurement. Mean assessment time was significantly shorter using the smartphone AI tool (1.8 ± 0.4 minutes) compared with ruler-based assessment (4.5 ± 0.8 minutes).
Conclusion Smartphone-based AI measurement provides accurate and rapid quantification of ptosis severity compared with traditional ruler assessment. This approach may offer a simple and accessible tool for objective eyelid measurement in routine oculoplastic practice.